The INVALCO Model 4528 EZ Detector Card is a micro processor based instrument. When used with an INVALCO Model CX-645 Capacitance Probe it is a continuous on line monitor and alarm of the percent BS&W in a flowing crude oil emulsion stream. When used in conjunction with an INVALCO Model 4728 BS&W Monitor, the user can obtain a real time visual indication of percent BS&W in a flowing crude oil stream.
Features
-
Field Selectable 0-5, 0-10, 0-20% BS&W detection range.
-
Button entry of cut, calibration, automatic span calibration, compensated for Dk.
-
Continuous detection and automatic indication of probe faults, and readings outside the calibrated range switchable fail-safe high or low.
-
A single red LED used to indicate various modes of operation.
Applications
LACT (Lease Automatic Custody Transfer) Units
Detect and reroute oil that has been contaminated by excess S&W.
Pipeline Loading
Monitor transfer of petroleum/condensate products from offshore loading facilities.
Dehydration Equipment
Determine and enhance equipment efficiencies.
Fuel Oil Monitoring
Determine contamination of fuel oil by condensation or other external factors before entry to engine.
Storage and Treating Facilities
Monitoring and early detection of undesirable conditions as well as interface detection during de-watering of storage tanks.
Power Supply
8-28 Vdc @ 130 mA
BS&W Range
0-5%, 0-10%, and 0-20%
Field Selectable
Capacitance Range
122 pf to 380 pf (approx.)
Dk Range
1.5 to 3.5
Detector Operating Voltage Range (Excl. 4-20mA)
11 to 28 VDC
4-20mA Mode
Loop powerable 4-20mA
Voltage Range - 8 to 28 V
Zero Fine Adjust - 3.75 to 4.25
Span 20 ± 0.5mA at 5.0 V